首页 | 本学科首页   官方微博 | 高级检索  
   检索      


LED-transistor optical vascular measure: Replicability of a new device
Authors:J Richard Jennings  Larry Young  Kay Brock
Institution:1. University of Pittsburgh, Pittsburgh, Pennsylvania
Abstract:We describe the use of a photo-optical device for the measurement of peripheral vascular activity, the Optek OPB707A, which we advocate to replace the now obsolete Sensor Technology STRT-850A. The Optek device provides a close coupling of light emitted and received through use of a GaAlAs light-emitting diode (LED). It also provides a photodarlington transistor as the receiver. The device should be used in a transmitted mode (e.g., LED illuminating tissue on volar surface of thumb and receiver on the thumbnail). Signal-to-noise characteristics and replicability of the signal from occasion to occasion are shown to be equal or superior to those of the earlier device.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏