A microcomputer-controlled solid-state dark adaptometer |
| |
Authors: | Peter A O’mara Harry Zwick Charles W Van Sice |
| |
Institution: | 1. Division of Ocular Hazards, Letterman Army Institute of Research, Presidio of San Francisco, 94129, California
|
| |
Abstract: | A low-cost microprocessor-controlled light-emitting diode (LED) dark adaptometer is described. The apparent intensities of red and green stimuli are controlled by changing the duty cycles of LED display elements which are operated at constant pulse repetition rates. The psychophysical method of limits is used to obtain threshold data. Stimulus parameters and test procedures are under software control. The design utilizes programmable integrated circuitry that may be used with a variety of microprocessors. |
| |
Keywords: | |